Calibration Report FlowMeter

A Calibration Report FlowMeter is a instance of a Flow Meter Device on the calibration report containing the information regarding the meter calibration. The flow meter variables (Frequency/Count, Differential Pressure, Static Pressure, Temperature, Water Cut and Density) are independent reports that are attached to the flow meter calibration report. This allows each variable to be viewed as a independent report.

Attributes

CalibrationReportFlowMeter_New

Process Variables

  • As Found - Live Flow Reading - The live flow rate reading of the device when left.
  • As Left - Live Flow Reading - The live flow rate reading of the device when found.
  • Public Notes - Available for any notes related to the flow meter calibration. Notes are public and visible to all Employees and Clients.
  • Calibration Acceptable - The flow meter calibration was completed successfully.

Run Data

  • Orifice Size - Actual - The actual orifice plate size from the plate itself. (Only if meter is a Orifice Plate type)
  • Orifice Size - As Found - The orifice plate size entered in the flow compter as found. (Only if meter is a Orifice Plate type)
  • Orifice Size - As Left - The orifice plate size entered in the flow compter as left. (Only if meter is a Orifice Plate type)
  • Pipe Size - Actual - The actual pipe size from the pipe itself.
  • Pipe Size - As Found - The pipe size entered in the flow compter as found.
  • Pipe Size - As Left - The pipe size entered in the flow compter as left.
  • Beta Ratio - The ration of orifice plate size to pipe size. If left blank this value will be automaticaaly calculated based on the sizes entered.
  • Public Notes - Available for any notes related to the flow meter calibration. Notes are public and visible to all Employees and Clients.
  • Calibration Acceptable - The meter calibration was completed successfully.

Configuration

  • Low Flow Cutoff - The differential pressure/frequency reading of the meter when the flow calculation will be stopped.
  • Base Pressure - The base pressure of the flow calculation.
  • Base Temperature - The base temperature of the flow calculation.
  • Tap Location - The location of the static pressure tap (Upstream/Downstream).
  • Static Pressure Measured in - If the static pressure is measured in gauge or absolute pressure.
  • Atmospheric Pressure is - If the atmospheric pressure is calculated or entered.
  • Atmospheric Pressure - The atmospheric pressure.
  • Heating Value is - If the heating value is calculated or entered.
  • Heating Value - The heating value.
  • Specific Gravity is - If the specific gravity is calculated or entered.
  • Specific Gravity - The specific gravity.
  • Compressibility Calculation - Select the compressibility calculation method used.
  • Gas Composition is - If the gas composition is libe from a gas chromatograph or manually entered.
  • Flow Calculation - Select the flow calculation method used.

Test data can also be added to the report by clicking the 'Add Test' button. Doing so will display a form to collect test data. Any number of tests can be added to the report.

Frequency/Count

  • As Found - Live Reading - The live frequency/count reading of the device when left.
  • As Left - Live Reading - The live frequency/count reading of the device when found.
  • Calibration Range - Low - The low range of the flow meter differential frequency/count calibration. This field will be automatically populated from the meter configuration.
  • Calibration Range - High - The high range of the flow meter differential frequency/count calibration. This field will be automatically populated from the meter configuration.
  • Public Notes - Available for any notes related to the device calibration. Notes are public and visible to all Employees and Clients.
  • Calibration Acceptable - The device calibration was completed successfully.

Frequency/Count Tests

  • Applied - The calibration value applied to the device.
  • Rising/Falling - Whether the trip was recorded on rising or falling value.
  • Reading as Found - The device reading at the applied value when found.
  • Reading as Left - The device reading at the applied value when left.
  • Percent Span - The percentage of range that was tested. If left blank this value will be automatically calculated based on the calibration range.

Differential Pressure

  • As Found - Live Reading - The live differential pressure reading of the device when left.
  • As Left - Live Reading - The live differential pressure reading of the device when found.
  • Calibration Range - Low - The low range of the flow meter differential pressure calibration. This field will be automatically populated from the meter configuration.
  • Calibration Range - High - The high range of the flow meter differential pressure calibration. This field will be automatically populated from the meter configuration.
  • Public Notes - Available for any notes related to the device calibration. Notes are public and visible to all Employees and Clients.
  • Calibration Acceptable - The device calibration was completed successfully.

Differential Pressure Tests

  • Applied - The calibration value applied to the device.
  • Rising/Falling - Whether the trip was recorded on rising or falling value.
  • Reading as Found - The device reading at the applied value when found.
  • Reading as Left - The device reading at the applied value when left.
  • Percent Span - The percentage of range that was tested. If left blank this value will be automatically calculated based on the calibration range.

Static Pressure

  • As Found - Live Reading - The live static pressure reading of the device when left.
  • As Left - Live Reading - The live static pressure reading of the device when found.
  • Calibration Range - Low - The low range of the flow meter static pressure calibration. This field will be automatically populated from the meter configuration.
  • Calibration Range - High - The high range of the flow meter static pressure calibration. This field will be automatically populated from the meter configuration.
  • Public Notes - Available for any notes related to the device calibration. Notes are public and visible to all Employees and Clients.
  • Calibration Acceptable - The device calibration was completed successfully.

Static Pressure Tests

  • Applied - The calibration value applied to the device.
  • Rising/Falling - Whether the trip was recorded on rising or falling value.
  • Reading as Found - The device reading at the applied value when found.
  • Reading as Left - The device reading at the applied value when left.
  • Percent Span - The percentage of range that was tested. If left blank this value will be automatically calculated based on the calibration range.

Temperature

  • As Found - Live Reading - The live temperature reading of the device when left.
  • As Left - Live Reading - The live temperature reading of the device when found.
  • Calibration Range - Low - The low range of the flow meter temperature calibration. This field will be automatically populated from the meter configuration.
  • Calibration Range - High - The high range of the flow meter temperature calibration. This field will be automatically populated from the meter configuration.
  • Public Notes - Available for any notes related to the device calibration. Notes are public and visible to all Employees and Clients.
  • Calibration Acceptable - The device calibration was completed successfully.

Temperature Tests

  • Applied - The calibration value applied to the device.
  • Rising/Falling - Whether the trip was recorded on rising or falling value.
  • Reading as Found - The device reading at the applied value when found.
  • Reading as Left - The device reading at the applied value when left.
  • Percent Span - The percentage of range that was tested. If left blank this value will be automatically calculated based on the calibration range.

Density (Liquid Meters Only)

  • As Found - Live Reading - The live density reading of the device when left.
  • As Left - Live Reading - The live density reading of the device when found.
  • Calibration Range - Low - The low range of the flow meter density calibration. This field will be automatically populated from the meter configuration.
  • Calibration Range - High - The high range of the flow meter density calibration. This field will be automatically populated from the meter configuration.
  • Public Notes - Available for any notes related to the device calibration. Notes are public and visible to all Employees and Clients.
  • Calibration Acceptable - The device calibration was completed successfully.

Density Tests

  • Applied - The calibration value applied to the device.
  • Rising/Falling - Whether the trip was recorded on rising or falling value.
  • Reading as Found - The device reading at the applied value when found.
  • Reading as Left - The device reading at the applied value when left.
  • Percent Span - The percentage of range that was tested. If left blank this value will be automatically calculated based on the calibration range.

Water Cut (Liquid Meters Only)

  • As Found - Live Reading - The live water cut reading of the device when left.
  • As Left - Live Reading - The live water cut reading of the device when found.
  • Calibration Range - Low - The low range of the flow meter water cut calibration. This field will be automatically populated from the meter configuration.
  • Calibration Range - High - The high range of the flow meter water cut calibration. This field will be automatically populated from the meter configuration.
  • Public Notes - Available for any notes related to the device calibration. Notes are public and visible to all Employees and Clients.
  • Calibration Acceptable - The device calibration was completed successfully.

Water Cut Tests

  • Applied - The calibration value applied to the device.
  • Rising/Falling - Whether the trip was recorded on rising or falling value.

  • Reading as Found - The device reading at the applied value when found.

  • Reading as Left - The device reading at the applied value when left.
  • Percent Span - The percentage of range that was tested. If left blank this value will be automaticaaly calculated based on the calibration range.

Composition

Composition data can also be added to the report by clicking the 'Add Composition' button. Doing so will display a form to collect composition data. Any number of compositions can be added to the report.

  • Type - The type of composition. (As Left/As Found/Verify).
  • Nitrogen (N2) - The nitrogen component. (mol %)
  • Carbon Dioxide (CO2) - The carbon dioxide component. (mol %)
  • Methane (C1) - The methane component. (mol %)
  • Ethane (C2) - The ethane component. (mol %)
  • Propane (C3) - The propane component. (mol %)
  • n-Butane (nC4) - The n-butane component. (mol %)
  • i-Butane (iC4) - The i-butane component. (mol %)
  • n-Pentane (nC5) - The n-pentane component. (mol %)
  • i-Pentane (iC5) - The i-pentane component. (mol %)
  • Hexane (C6) - The hexane component. (mol %)
  • Heptane (C7) - The heptane component. (mol %)
  • Octane (C8) - The octane component. (mol %)
  • Nonane (C9) - The nonane component. (mol %)
  • Decane (C10) - The decane component. (mol %)
  • Hydrogen Sulphide (H2S) - The hydrogen sulphide component. (mol %)
  • Water (H2O) - The water component. (mol %)
  • Helium (He) - The helium component. (mol %)
  • Oxygen (O2) - The oxygen component. (mol %)
  • Carbon Monoxide (CO) - the carbon monoxide component. (mol %)
  • Hydrogen (H2) - The hydrogen component. (mol %)
  • Argon (Ar) - The argon component. (mol %)
  • Composition Total - The composition total. If left blank this value will be automatically calculated based on the entered values.

Flow Tests

  • Test Flow - Applied test flow rate.
  • Measured Flow - Flow meter measured test flow rate.
  • Measured Frequency - Primary element measured output frequency (linear meters only).
  • k-Factor - Primary element k-factor (linear meters only).
  • m-Factor - Primary element m-factor (linear meters only).

Employee Access

Employee List View Detail View Create New Edit Existing Delete Accessible Meter Reports
Administrator Yes Yes Yes Yes Yes All meters belonging to calibration reports of the company
Head Office Manager Yes Yes Yes Yes Yes All meters belonging to calibration reports of the company
Head Office Administrator Yes Yes Yes Yes Yes All meters belonging to calibration reports of the company
Branch Office Manager Yes Yes Yes Yes Yes All meters belonging to calibration reports of the branch user is associated with
Branch Office Administrator Yes Yes Yes Yes Yes All meters belonging to calibration reports of the branch user is associated with
Employee Yes Yes Yes Yes Yes All meters belonging to calibration reports of the branch user is associated with

Client Access (when service provider enabled)

Client List View Detail View Create New Edit Existing Delete Accessible Meter Reports
Customer Manager Yes Yes No No No All meters belonging to calibration reports of the customer
Customer Operator Yes Yes No No No All meters belonging to calibration reports of the customer
Area Manager Yes Yes No No No All meters belonging to calibration reports in permitted areas
Area Operator Yes Yes No No No All meters belonging to calibration reports in permitted areas
Location Manager Yes Yes No No No All meters belonging to calibration reports in permitted locations
Location Operator Yes Yes No No No All meters belonging to calibration reports in permitted locations

List Views

CalibrationReportFlowMeter_List

The Calibration Report Flow Meters list view is available from the CalibrationReport. Clicking the 'Add Flow Meter' button at the top of the view will display the list view. The list view displays the Flow Meter Devices available at the location for calibration and clicking the tab 'Devices On Report' will display the devices currently on the Calibration Report.

The lists are sorted by meter tag.

Details

CalibrationReportFlowMeter_Show

The Calibration Report Flow Meter detail view displays a full report of the calibration data recorded during the calibration.